Select a Country or Region

22
22

Low Flux Hemodialyser

  • Features
  • High capability of toxic clearance.
  • Excellent biocompatibility.
  • Higher performance of small and medium size removal.
  • Lower loss of Albumin.

Contact US

Share to

In Vitro performance data / technical data

 

  A40 A60 A80 A200
Ultrafitration Coeff.(mL/h x mmHg) 18 20 22 24
Clearance QB =200mL/min  
Urea 190 192 194 196
Creatinine 181 184 187 190
Phosphate 175 180 185 189
Vitamin B12 110 118 126 133
Clearance QB =300mL/min  
Urea 252 260 267 275
Creatinine 233 245 257 270
Phosphate 214 225 235 245
Vitamin B12 120 136 150 165
Clearance QB=400mL/min  
Urea 279 290 300 312
Creatinine 261 275 288 300
Phosphate 244 257 270 283
Vitamin B12 125 142 158 173

 

Remarks: In Vitro performance: QD= 500mL/min, QF=10mL/min, T=37℃ (EN1283, ISO8637)
Ultrafi ltration coeffi cients: Bovine plasma, TP=60g/L ± 5g/L, TMP=10mm Hg,

 

 

Effective Surface(㎡) 1.4 1.6 1.8 2
Blood flow range(ml/min) 200-400 200-400 200-400 200-400
Wall thickness/lumen(um) 40/200 40/200 40/200 40/200
Priming volume(mL) 78 86 99 108
Membrane material Polyethersulfone or polysulfone
Maximum TMP(mm Hg) 500 500 500 500
Housing material polycarbonate
Potting compound Polyurethane
Cover material polycarbonate
O-ring Silicone Rubber
Caps Polypropylene
Sterilisation method Gamma
Form of treatment HD
Units per box 24
Product code A-40 A-60 A-80 A-200

Contact us

+011-86-23-88192230
Century Emperor North Tower, Jiangbei District, Chongqing,China

About Dezhixing